device · product 1 of 1
Philips Laser System, REF: LAS-100. Used in minimally invasive interventional procedures within the cardiovascular system, and for the removal of problematic pacemaker and defibrillator cardiac leads.
- Recall number
- Z-0221-2023
- Initiated
- October 14, 2022
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Spectranetics Corporation
- Quantity
- 113
App-derived interpretation
Laser System may detect Error 106 or Error 108 System Failures that will cause the system to enter non-recoverable safe state. The user will not be able to proceed until the error is cleared. May result in initiation of treatment delay and/or treatment beyond initial scope and/or unable to treat patient.
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
Laser System may detect Error 106 or Error 108 System Failures that will cause the system to enter non-recoverable safe state. The user will not be able to proceed until the error is cleared. May result in initiation of treatment delay and/or treatment beyond initial scope and/or unable to treat patient.
Code information
UDI-00813132027452, Software Version: 0.5.0.3, Serial Numbers: 100000, 100001, 100002, 100003, 100004, 100005, 100006, 100007, 100008, 100009, 100010, 100011, 100012, 100013, 100014, 100015, 100016, 100017, 100018, 100019, 100020, 100021, 100022, 100023, 100024, 100025, 100026, 100027, 100028, 100029, 100030, 100033, 100034, 100035, 100036, 100037, 100038, 100039, 100040, 100041, 100042, 100043, 100044, 100045, 100046, 100047, 100048, 100049, 100050, 100051, 100052, 100053, 100054, 100055, 100058, 100059, 100060, 100061, 100062, 100063, 100064, 100065, 100066, 100067, 100068, 100069, 100070, 100071, 100072, 100073, 100074, 100075, 100076, 100077, 100078, 100079, 100080, 100081, 100082, 100083, 100084, 100085, 100086, 100087, 100088, 100089, 100090, 100091, 100092, 100093, 100094, 100095, 100096, 100097, 100098, 100099, 100100, 100101, 100102, 100103, 100104, 100105, 100106, 100107, 100108, 100112, 100113, 100114, 100115, 100116, 100117, 100118, 100120
Distribution pattern
US: LA, MO, TN, CA, NC, PA, HI, AZ, CO, OK, FL, TX, MN, MI, VA, CT, NY, MD, IL, NJ, MA, KS, ME, AR, IA, IN, GA, ID, OR. OUS: DE, GB, ES, SE, IT, RS, GR, IN, FR, IL, DK, SA