Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 91104

9 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 20, 2022
Product types
Device
Classifications
Class I
Statuses
Terminated
Recalling firm wording
Remel Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

9 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 9

Thermo SCIENTIFIC Sensititre Plate, GN4F, Gram Negative IVD AST

Z-0327-2023
Recall number
Z-0327-2023
Initiated
October 20, 2022
Classification
Class I
Status
Terminated
Recalling firm
Remel Inc
Quantity
1917 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.

Code information

Lot Numbers: B2232B, B2197, B2012, B1355, B1181, B1173, B1125A

Distribution pattern

US Nationwide Distribution

device · product 2 of 9

Thermo SCIENTIFIC, Sensititre GN6F, Gram Negative IVD AST

Z-0328-2023
Recall number
Z-0328-2023
Initiated
October 20, 2022
Classification
Class I
Status
Terminated
Recalling firm
Remel Inc
Quantity
2471 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.

Code information

Lot Numbers: B2247, B2096, B1445, B1263, B1176, B0503A

Distribution pattern

US Nationwide Distribution

device · product 3 of 9

Thermo SCIENTIFIC, Sensititre GN7F, Gram Negative IVD AST

Z-0329-2023
Recall number
Z-0329-2023
Initiated
October 20, 2022
Classification
Class I
Status
Terminated
Recalling firm
Remel Inc
Quantity
8073 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.

Code information

Lot Numbers: B2317, B2267, B2251, B2177, B2086, B2037, B2034B, B1476, B1424, B1384, B1286, B1215, B1182, B1104, B1051, B1021

Distribution pattern

US Nationwide Distribution

device · product 4 of 9

Thermo SCIENTIFIC, Sensititre HPB1, Gram Negative IVD AST

Z-0330-2023
Recall number
Z-0330-2023
Initiated
October 20, 2022
Classification
Class I
Status
Terminated
Recalling firm
Remel Inc
Quantity
665 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.

Code information

Lot Numbers: B2206A, B2035A, B1333, B1144A, B0483

Distribution pattern

US Nationwide Distribution

device · product 5 of 9

Thermo SCIENTIFIC, Sensititre CMC5VGNF, Gram Negative IVD AST

Z-0331-2023
Recall number
Z-0331-2023
Initiated
October 20, 2022
Classification
Class I
Status
Terminated
Recalling firm
Remel Inc
Quantity
588 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.

Code information

Lot Numbers: B2125, B2115

Distribution pattern

US Nationwide Distribution

device · product 6 of 9

Thermo SCIENTIFIC, Sensititre CMC7AFLF, Gram Negative IVD AST

Z-0332-2023
Recall number
Z-0332-2023
Initiated
October 20, 2022
Classification
Class I
Status
Terminated
Recalling firm
Remel Inc
Quantity
1647 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.

Code information

Lot Numbers: B2281, B2151A, B2034, B1453

Distribution pattern

US Nationwide Distribution

device · product 7 of 9

Thermo SCIENTIFIC, Sensititre MDRGN3F, Gram Negative IVD AST

Z-0333-2023
Recall number
Z-0333-2023
Initiated
October 20, 2022
Classification
Class I
Status
Terminated
Recalling firm
Remel Inc
Quantity
70 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.

Code information

Lot Numbers: B2311B

Distribution pattern

US Nationwide Distribution

device · product 8 of 9

Thermo SCIENTIFIC, Sensititre STP6F, Gram Negative IVD AST

Z-0334-2023
Recall number
Z-0334-2023
Initiated
October 20, 2022
Classification
Class I
Status
Terminated
Recalling firm
Remel Inc
Quantity
2656 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.

Code information

Lot Numbers: B2314, B2174, B2036, B1346, B1183, B1115, B1027

Distribution pattern

US Nationwide Distribution

device · product 9 of 9

Thermo SCIENTIFIC, Sensititre STP7F, Gram Negative IVD AST

Z-0335-2023
Recall number
Z-0335-2023
Initiated
October 20, 2022
Classification
Class I
Status
Terminated
Recalling firm
Remel Inc
Quantity
9 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.

Code information

Lot Numbers: B2265A

Distribution pattern

US Nationwide Distribution