Open FDA event record

Event summary

Timeline bucket
December 09, 2022
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Abbott Molecular, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Alinity m System. The Alinity m System is designed to automate the steps for nucleic acid testing, which includes sample processing, amplification, detection, and data reduction.

Z-1009-2023
Recall number
Z-1009-2023
Initiated
December 09, 2022
Classification
Class II
Status
Ongoing
Recalling firm
Abbott Molecular, Inc.
Quantity
918 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The firm identified the following three potential issues which may have impacted ADUs installed on distributed Alinity m Systems. 1. Systems which had their ADUs serviced may have an incorrect board value (calibration data file) referenced causing a misconfiguration of the calibration. 2. Inability of Maintenance and Diagnostics 2300 (Amp Detect Optical Calibrations) to recognize error handling and thermal hold errors during the calibration process, potentially leading to a misconfiguration of the calibration. 3. Under specific conditions during the ADU manufacturing process, an incorrect calibration may potentially have been configured.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The firm identified the following three potential issues which may have impacted ADUs installed on distributed Alinity m Systems. 1. Systems which had their ADUs serviced may have an incorrect board value (calibration data file) referenced causing a misconfiguration of the calibration. 2. Inability of Maintenance and Diagnostics 2300 (Amp Detect Optical Calibrations) to recognize error handling and thermal hold errors during the calibration process, potentially leading to a misconfiguration of the calibration. 3. Under specific conditions during the ADU manufacturing process, an incorrect calibration may potentially have been configured.

Code information

Model Number: 08N53-002