device · product 1 of 1
Alinity m System. The Alinity m System is designed to automate the steps for nucleic acid testing, which includes sample processing, amplification, detection, and data reduction.
- Recall number
- Z-1009-2023
- Initiated
- December 09, 2022
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Abbott Molecular, Inc.
- Quantity
- 918 units
App-derived interpretation
The firm identified the following three potential issues which may have impacted ADUs installed on distributed Alinity m Systems. 1. Systems which had their ADUs serviced may have an incorrect board value (calibration data file) referenced causing a misconfiguration of the calibration. 2. Inability of Maintenance and Diagnostics 2300 (Amp Detect Optical Calibrations) to recognize error handling and thermal hold errors during the calibration process, potentially leading to a misconfiguration of the calibration. 3. Under specific conditions during the ADU manufacturing process, an incorrect calibration may potentially have been configured.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Inspect official wording and provenance
Reason for recall
The firm identified the following three potential issues which may have impacted ADUs installed on distributed Alinity m Systems. 1. Systems which had their ADUs serviced may have an incorrect board value (calibration data file) referenced causing a misconfiguration of the calibration. 2. Inability of Maintenance and Diagnostics 2300 (Amp Detect Optical Calibrations) to recognize error handling and thermal hold errors during the calibration process, potentially leading to a misconfiguration of the calibration. 3. Under specific conditions during the ADU manufacturing process, an incorrect calibration may potentially have been configured.
Code information
Model Number: 08N53-002