Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 91290

3 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 07, 2022
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Johnson & Johnson Surgical Vision, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

3 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 3

Veritas Advanced Infusion Packs, REF: VRT-AI

Z-1005-2023
Recall number
Z-1005-2023
Initiated
December 07, 2022
Classification
Class II
Status
Ongoing
Quantity
3133

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

Code information

UDI-DI/Lots: 05050474700901/60400378, 60401981, 60401982

Distribution pattern

US: FL, MS, NY, TN, TX, IL, MN, OH. OUS: Belgium, China, Estonia, France, Germany, Indonesia, Italy, Japan, Malaysia, Netherlands, Portugal, Sri Lanka, Switzerland, Thailand, United Kingdom

device · product 2 of 3

FUSION Dual Pump Packs, REF: OPO73

Z-1006-2023
Recall number
Z-1006-2023
Initiated
December 07, 2022
Classification
Class II
Status
Ongoing
Quantity
7644

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

Code information

UDI-DI/Lots: 15050474602080/60373238, 60394037, 60308197

Distribution pattern

US: FL, MS, NY, TN, TX, IL, MN, OH. OUS: Belgium, China, Estonia, France, Germany, Indonesia, Italy, Japan, Malaysia, Netherlands, Portugal, Sri Lanka, Switzerland, Thailand, United Kingdom

device · product 3 of 3

Single-Use Packs, REF: OPO80

Z-1007-2023
Recall number
Z-1007-2023
Initiated
December 07, 2022
Classification
Class II
Status
Ongoing
Quantity
384

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

Code information

UDI-DI/Lot: 05050474573352/60326033

Distribution pattern

US: FL, MS, NY, TN, TX, IL, MN, OH. OUS: Belgium, China, Estonia, France, Germany, Indonesia, Italy, Japan, Malaysia, Netherlands, Portugal, Sri Lanka, Switzerland, Thailand, United Kingdom