openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
device · product 1 of 1
Mammotome MammoStar Biopsy Site Identified, REF STAR1401
The label incorrectly identifies the product as STAR1401, a 1x3mm Barbell shaped marker pre-loaded in a 14 gauge needle, while the packaging contains STAR1402, a 1x5mm Tribell shaped marker pre-loaded in a 14 gauge needle.
These labels are deterministic app interpretations, not FDA categories.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
The label incorrectly identifies the product as STAR1401, a 1x3mm Barbell shaped marker pre-loaded in a 14 gauge needle, while the packaging contains STAR1402, a 1x5mm Tribell shaped marker pre-loaded in a 14 gauge needle.