Event summary

Timeline bucket
January 24, 2023
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
CareFusion 303, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Alaris PC Unit 8015 utilizing 802.11 b/g or 802.11 a/b/g wireless network cards, with IP addresses from x.x.x.224 to x.x.x.255

Z-1192-2023
Recall number
Z-1192-2023
Initiated
January 24, 2023
Classification
Class II
Status
Ongoing
Recalling firm
CareFusion 303, Inc.
Quantity
83,185

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software/network cards/IP addresses can have network/wireless connection failure. Could cause

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software Design Change

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Infusion pump PCs with specific software/network cards/IP addresses can have network/wireless connection failure. Could cause inability to receive new datasets and inability to transfer retrospective analytics data to connected systems. Users with interfaces to electronic medication record or admin discharge transfer, may be unable to receive/send data; infusions may need to be manually programmed

Code information

All PC Units, UDI-DI: 10885403801549, 10885403801532, with software versions 12.1.0, 12.1.2, 12.1.3, utilizing affected network cards and IP addresses

Distribution pattern

US: FL, CA, OR, IL, WI, NJ, AZ, NV, CO, WY, NE, TX, MI, VA, OH, NC, LA, NM, AL, TN, AR, PA, CT, IN, KS, GA, SC, MD, UT, MA, DC, OK, IA, NY, WA, SD, ND, MN, KY, MO, WV, MS, DE, ME, MT, ID, NH, HI, AK, VT OUS: Canada, Qatar, Saudi Arabia,, Kuwait, United Arab Emirates, Israel, South Africa, Zimbabwe, Botswana, Australia, New Zealand