device · product 1 of 1
Tri-Staple 2.0 Black Intelligent Reload, Product Number SIG60AXT
- Recall number
- Z-1228-2023
- Initiated
- January 25, 2023
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Covidien, LP
- Quantity
- 264 US; 5056 OUS
App-derived interpretation
Affected lots have the potential for a broken sled vane, which may cause the reload to misfire leading to non-functional staple line closure, transecting tissue without forming staples, and tissue hang-up. These conditions may be associated to a delay to treatment, unspecified infection, hemorrhage/blood loss/bleeding, failure to anastomose, peritonitis, sepsis, pneumothorax, tissue trauma, or death.
Official device-enrichment evidence · Sourced
Process change control
Inspect official wording and provenance
Reason for recall
Affected lots have the potential for a broken sled vane, which may cause the reload to misfire leading to non-functional staple line closure, transecting tissue without forming staples, and tissue hang-up. These conditions may be associated to a delay to treatment, unspecified infection, hemorrhage/blood loss/bleeding, failure to anastomose, peritonitis, sepsis, pneumothorax, tissue trauma, or death.
Code information
UDI-DI (GTIN): 20884521543598 Lots: N2D0002Y, N2D0004Y, N2D0195Y
Distribution pattern
Worldwide distribution - US Nationwide distribution in the states of Arkansas, California, Florida, Georgia, Louisiana, Massachusetts, New York, North Carolina, Ohio, Washington and the countries of Australia, Austria, Belgium, Canada, Canary Islands, Denmark, Finland, France, Germany, Ireland, Italy, Kuwait, Netherlands, New Caledonia, Norway, Poland, Portugal, Spain, Sweden, Switzerland, United Kingdom.