Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 91666

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 07, 2023
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Stryker GmbH

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

stryker Hoffmann LRF, Transport Strut, Model Number 4933-0-400; Orthopedic external fixation device intended to keep bone fragments/segments in a correct anatomical position until reliable bone consolidation is achieved

Z-1191-2023
Recall number
Z-1191-2023
Initiated
February 07, 2023
Classification
Class II
Status
Ongoing
Recalling firm
Stryker GmbH
Quantity
488 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Stryker has identified a nonconformance in specific lots of Hoffmann LRF Bone Transport Struts. Specifically, the thread pitch is oversized at the distal end of the threaded rod, and as a result the distal ring fixation cannot be locked between 0- 50mm of the threaded rod.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Stryker has identified a nonconformance in specific lots of Hoffmann LRF Bone Transport Struts. Specifically, the thread pitch is oversized at the distal end of the threaded rod, and as a result the distal ring fixation cannot be locked between 0- 50mm of the threaded rod.

Code information

Serial numbers: D32158, D32160, D32161, D32162, D32163, D32164, D34343, D34344, D40690, D41629, D41630, D41631, D44713, G16331, G16332, G16333, G26278, G29759, G29760, G32730, G32731, G32735, G45119, H30398, H30399, H66392

Distribution pattern

US Nationwide distribution.