Habib EndoHPB Bipolar Radiofrequency (RF) Catheter 8F (2.7mm). A radiofrequency (RF) catheter which provides bipolar energy to perform partial or complete ablation of tissue in the pancreatic and biliary tracts. UPN: M00500070
- Recall number
- Z-1397-2023
- Initiated
- February 02, 2023
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Boston Scientific Corporation
- Quantity
- 6635 units US: 1815 US units; 4820 OUS units
App-derived interpretation
Cease use of the Habib EndoHPB Bipolar RF Catheter with the ERBE VIO3 Generator immediately due to the potential for delivery of excessive energy leading to thermal injury or tissue damage. An inaccurate setting is currently provided in the HABIB EndoHPB Catheter's Instructions for Use ("IFU").
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Inspect official wording and provenance
Reason for recall
Cease use of the Habib EndoHPB Bipolar RF Catheter with the ERBE VIO3 Generator immediately due to the potential for delivery of excessive energy leading to thermal injury or tissue damage. An inaccurate setting is currently provided in the HABIB EndoHPB Catheter's Instructions for Use ("IFU").
Code information
GTIN: 8714729981909 All lots , All Expiration Dates
Distribution pattern
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Bulgaria, Canada, Costa Rica, Czech Republic, Egypt, Finland, France, Germany, Great Britain, Greece, Hong Kong, Hungary, Iceland, Iraq, Israel, Italy, Japan, Korea, Kuwait, Latvia, Lithuania, Mexico, Netherlands, New Zealand, Oman, Panama, Poland, Portugal, Romania, Russian Fed., Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Unit. Arab Emir.