Bard Mission Disposable Core Biopsy Instrument Kit: REF: 2010MSK, 20G x 10CM; REF: 2016MSK, 20G x 16CM
- Recall number
- Z-1564-2023
- Initiated
- April 11, 2023
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Bard Peripheral Vascular Inc
- Quantity
- 4,120
App-derived interpretation
Coaxial cannula internal diameter and length of biopsy instruments exceed the external diameter and length of the biopsy needle, prohibiting the biopsy needle from properly fitting into the coaxial cannula and accessing the target tissue, which could lead to procedure delay, and insufficient sample acquisition requiring a repeat procedure.
Official device-enrichment evidence · Sourced
Mixed-up of materials/components
Inspect official wording and provenance
Reason for recall
Coaxial cannula internal diameter and length of biopsy instruments exceed the external diameter and length of the biopsy needle, prohibiting the biopsy needle from properly fitting into the coaxial cannula and accessing the target tissue, which could lead to procedure delay, and insufficient sample acquisition requiring a repeat procedure.
Code information
REF/UDI-DI/Lot (Expiration Date): 2010MSK/00801741097089/0001445946(11/28/2024), 2016MSK/00801741097096/1458844(02/28/2025)
Distribution pattern
US Nationwide Distribution: PR, AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY . OUS: BE, CA, AU