device · product 1 of 1
Ultraview SL (UVSL) Command Module, Model 91496
- Recall number
- Z-1503-2023
- Initiated
- April 12, 2023
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Spacelabs Healthcare, Inc.
- Quantity
- 1,088
App-derived interpretation
software version: 1) Older versions will
Official device-enrichment evidence · Sourced
Component change control
Inspect official wording and provenance
Reason for recall
Command modules will display the following when parameter processing is suspended, depending on software version: 1) Older versions will have waveforms present with no alarms, 2) Newer versions have waveforms absent with no alarms. This may cause clinicians to miss a potentially catastrophic event.
Code information
Software version 2.03.06 and earlier and software version 2.04 and later.
Distribution pattern
US Nationwide distribution in the states of SD, KY, MO, WA, OH, NC, CA, OR, TX, GA, IL, OK, SC, PA, AL, KS, FL, ID, MI, MN, IA, NJ, MD, VA, AR, CT, NE, LA, AK, MS, ME, WV, NY, AZ, TN, WY, IN, MA, MT, CO, DE, ND, WI, NV, RI, NH, UT.