Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 91957

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
March 17, 2023
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Medtronic Neuromodulation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Medtronic SenSight Extension Tunneler Kit, REF B31030; The Extension Tunneler Kit is used to assist in the tunneling and passage of extensions for deep brain stimulation

Z-1408-2023
Recall number
Z-1408-2023
Initiated
March 17, 2023
Classification
Class II
Status
Ongoing
Recalling firm
Medtronic Neuromodulation
Quantity
1414 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Medtronic has received reports of SenSight Extension Tunneler kits that contain dual carriers with have been machined on only one side. As a result, the dual carrier cannot be used to pass two extensions simultaneously.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Medtronic has received reports of SenSight Extension Tunneler kits that contain dual carriers with have been machined on only one side. As a result, the dual carrier cannot be used to pass two extensions simultaneously.

Code information

UDI/DI00763000426361, Serial Numbers: HG56BR6, HG56C4R, HG56CJAH01, HG56CLE, HG586N7, HG586UM, HG58704, HG58VM0, HG58VQH, HG58VQHH01, HG58VY5, HG58W7H, HG58WWC, HG58WYB, HG58WYBH01, HG58X0E, HG597BH, HG597DH, HG597M8, HG59ABH, HG59AFC, HG59AV6, HG59RY0, HG5A9GT, HG5A9VY, HG5A9VZ, HG5AQH0, HG5AQQE, HG5BQ5Y, HG5BQLR, HG5BQPJ, HG5BTW7, HG5BTZX, HG5BU5W, HG5BU77, HG5EPSB, HG5EQ0C, HG5EQ2A, HG5EQ69, HG5F31N, HG5F338, HG5F3AZ, HG5F3EH, HG5F3G6, HG5F3K5, HG5F3NF, HG5FP0Y, HG5FP3B, HG5FP70, HG5FP70H01, HG5FPBE, HG5HQBM, HG5HQF8, HG5LJTL, HG5LJZQ, HG5LK2J, HG5LK4H, HG5LK5Y, HG5LK7C, HG5LK8P, HG5LK92, HG5LKAL, HG5LKC9, HG5LKEL, HG5LKGT, HG5LR02, HG5LR33, HG5LTWA, HG5LTWP, HG5LTXU, HG5LTYY, HG5LU12; UDI/DI 00763000486518, Serial Numbers: HG59ABH, HG59RY0, HG59SJ4, HG59SNC, HG5A9H3, HG5A9NB, HG5A9VY, HG5BQ5Y, HG5BQAW, HG5BQLR.

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canary Islands, Czech Republic, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Luxembourg, Netherlands, New Zealand, Portugal, Slovakia, Spain, Sweden, Switzerland, United Kingdom.