Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 92179

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
April 05, 2023
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Siemens Healthcare Diagnostics, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

ADVIA Chemistry ToxAmmonia Calibrator- In vitro diagnostic use in the calibration of ethanol (ETOH_2), ammonia (AMM), acetaminophen (ACET), and salicylate (SAL) assays on the ADVIA¿ Chemistry systems Siemens Material Number (SMN): 10309217

Z-1631-2023
Recall number
Z-1631-2023
Initiated
April 05, 2023
Classification
Class II
Status
Ongoing
Quantity
447 unit US; 1537 units OUS

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Reassignment of the Atellica CH Toxicology Calibrators (TOX CAL) and ADVIA Chemistry ToxAmmonia Calibrators for the Atellica CH Salicylate (Sal) Assay due to a positive bias

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Reassignment of the Atellica CH Toxicology Calibrators (TOX CAL) and ADVIA Chemistry ToxAmmonia Calibrators for the Atellica CH Salicylate (Sal) Assay due to a positive bias

Code information

UDI-DI: 00630414479286 Lot Numbers: 9520531 Exp.Date: 2023-05-31; 9520341 Exp. Date: 2024-03-31

Distribution pattern

US Nationwide and Worldwide Distribution Foreign: Argentina Australia Austria¿¿¿¿¿¿¿¿ Bahrain¿¿¿¿¿¿¿¿ Bangladesh Belgium¿¿¿¿¿¿¿¿ Bosnia¿Herzeg.¿ Brazil Canada Chile China Colombia Czech¿Republic¿ Egypt¿¿¿¿¿¿¿¿¿¿ Estonia¿¿¿¿¿¿¿¿ Finland¿¿¿¿¿¿¿¿ France¿¿¿¿¿¿¿¿¿ Georgia¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Greece¿¿¿¿¿¿¿¿¿ Hong Kong Hungary¿¿¿¿¿¿¿¿ India Israel¿¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Latvia¿¿¿¿¿¿¿¿¿ Malaysia Mexico Netherlands¿¿¿¿ Norway¿¿¿¿¿¿¿¿¿ Pakistan¿¿¿¿¿¿¿ Poland¿¿¿¿¿¿¿¿¿ Portugal¿¿¿¿¿¿¿ Republic of Korea Saudi¿Arabia¿¿¿ Serbia¿¿¿¿¿¿¿¿¿ Slovakia¿¿¿¿¿¿¿ South Africa Spain¿¿¿¿¿¿¿¿¿¿ Sweden¿¿¿¿¿¿¿¿¿ Taiwan Turkey¿¿¿¿¿¿¿¿¿ United¿Kingdom¿ Uruguay Vietnam

device · product 2 of 2

Atellica CH Toxicology Calibrator (TOX CAL)-The Atellica¿ CH Toxicology Calibrator (TOX CAL) is for in vitro diagnostic use in calibrating the Acet, ETOH and Sal assays using the Atellica¿ CH Analyzer Siemens Material Number (SMN): 11099440

Z-1632-2023
Recall number
Z-1632-2023
Initiated
April 05, 2023
Classification
Class II
Status
Ongoing
Quantity
6546 units US; 3681 units OUS

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Reassignment of the Atellica CH Toxicology Calibrators (TOX CAL) and ADVIA Chemistry ToxAmmonia Calibrators for the Atellica CH Salicylate (Sal) Assay due to a positive bias

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Reassignment of the Atellica CH Toxicology Calibrators (TOX CAL) and ADVIA Chemistry ToxAmmonia Calibrators for the Atellica CH Salicylate (Sal) Assay due to a positive bias

Code information

UDi-DI: 00630414597614 Lot Numbers: 74285306 Exp. Date: 2023-05-31; 74536672 Exp Date: 2024-03-31; 74732386 Exp. Date: 2024-11-30

Distribution pattern

US Nationwide and Worldwide Distribution Foreign: Argentina Australia Austria¿¿¿¿¿¿¿¿ Bahrain¿¿¿¿¿¿¿¿ Bangladesh Belgium¿¿¿¿¿¿¿¿ Bosnia¿Herzeg.¿ Brazil Canada Chile China Colombia Czech¿Republic¿ Egypt¿¿¿¿¿¿¿¿¿¿ Estonia¿¿¿¿¿¿¿¿ Finland¿¿¿¿¿¿¿¿ France¿¿¿¿¿¿¿¿¿ Georgia¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Greece¿¿¿¿¿¿¿¿¿ Hong Kong Hungary¿¿¿¿¿¿¿¿ India Israel¿¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Latvia¿¿¿¿¿¿¿¿¿ Malaysia Mexico Netherlands¿¿¿¿ Norway¿¿¿¿¿¿¿¿¿ Pakistan¿¿¿¿¿¿¿ Poland¿¿¿¿¿¿¿¿¿ Portugal¿¿¿¿¿¿¿ Republic of Korea Saudi¿Arabia¿¿¿ Serbia¿¿¿¿¿¿¿¿¿ Slovakia¿¿¿¿¿¿¿ South Africa Spain¿¿¿¿¿¿¿¿¿¿ Sweden¿¿¿¿¿¿¿¿¿ Taiwan Turkey¿¿¿¿¿¿¿¿¿ United¿Kingdom¿ Uruguay Vietnam