Open FDA event record

Event summary

Timeline bucket
April 25, 2023
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Davol, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

4 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 4

Ventralight ST 4" x 6" Ellipse with Echo PS Catalog Number: 5955460

Z-2241-2023
Recall number
Z-2241-2023
Initiated
April 25, 2023
Classification
Class II
Status
Ongoing
Recalling firm
Davol, Inc.
Quantity
1188 units US; 5 units OUS

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Balloon may not inflate as intended. If the balloon does not inflate during laparoscopic ventral hernia repair this may result in a delay or prolongation of the procedure, inadequate mesh placement, erosion/migration, pain and mesh infection

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Balloon may not inflate as intended. If the balloon does not inflate during laparoscopic ventral hernia repair this may result in a delay or prolongation of the procedure, inadequate mesh placement, erosion/migration, pain and mesh infection

Code information

UDI-DI: (01)00801741031724(17)240428(10) Lot Numbers: HUGS0481 HUGS0549 HUGS1019 HUGS2031 HUGS2079 HUGT1627 HUGS2025 HUGT1989

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Belgium, Czech Republic, France, Germany, Ireland, Italy, Netherlands, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK, India, Hong Kong, Greece, Chile, and Colombia.

device · product 2 of 4

Ventralight ST 4.5" Circle with Echo PS- A low profile, bioresorbable, coated, permanent mesh, with a pre-attached removable positioning system, designed for the reconstruction of soft tissue deficiencies during laparoscopic ventral hernia repair. Catalog Number: 5955450

Z-2242-2023
Recall number
Z-2242-2023
Initiated
April 25, 2023
Classification
Class II
Status
Ongoing
Recalling firm
Davol, Inc.
Quantity
1599 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Balloon may not inflate as intended. If the balloon does not inflate during laparoscopic ventral hernia repair this may result in a delay or prolongation of the procedure, inadequate mesh placement, erosion/migration, pain and mesh infection

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Balloon may not inflate as intended. If the balloon does not inflate during laparoscopic ventral hernia repair this may result in a delay or prolongation of the procedure, inadequate mesh placement, erosion/migration, pain and mesh infection

Code information

UDI-DI: (01)00801741031724(17)240428(10) Lot Numbers: Lot Number HUGS0922 HUGS2026 HUGS2061 HUGS2020 HUGS2044 HUGS2015 HUGS2076 HUGT1582 HUGU0750 HUGU0748

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Belgium, Czech Republic, France, Germany, Ireland, Italy, Netherlands, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK, India, Hong Kong, Greece, Chile, and Colombia.

device · product 3 of 4

Ventralight ST w Echo PS 4.5" Circle A low profile, bioresorbable, coated, permanent mesh, with a pre-attached removable positioning system, designed for the reconstruction of soft tissue deficiencies during laparoscopic ventral hernia repair. Catalog Number: 5955450G (EU only)

Z-2243-2023
Recall number
Z-2243-2023
Initiated
April 25, 2023
Classification
Class II
Status
Ongoing
Recalling firm
Davol, Inc.
Quantity
169 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Balloon may not inflate as intended. If the balloon does not inflate during laparoscopic ventral hernia repair this may result in a delay or prolongation of the procedure, inadequate mesh placement, erosion/migration, pain and mesh infection

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Balloon may not inflate as intended. If the balloon does not inflate during laparoscopic ventral hernia repair this may result in a delay or prolongation of the procedure, inadequate mesh placement, erosion/migration, pain and mesh infection

Code information

UDI-DI: (01)00801741201950(17)240528(10) Lot Number HUGT1609

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Belgium, Czech Republic, France, Germany, Ireland, Italy, Netherlands, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK, India, Hong Kong, Greece, Chile, and Colombia.

device · product 4 of 4

Ventralight ST w EchoPS 4"x6" Ellipse- A low profile, bioresorbable, coated, permanent mesh, with a pre-attached removable positioning system, designed for the reconstruction of soft tissue deficiencies during laparoscopic ventral hernia repair. Catalog Number: 5955460G (EU Only)

Z-2244-2023
Recall number
Z-2244-2023
Initiated
April 25, 2023
Classification
Class II
Status
Ongoing
Recalling firm
Davol, Inc.
Quantity
N/A

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Balloon may not inflate as intended. If the balloon does not inflate during laparoscopic ventral hernia repair this may result in a delay or prolongation of the procedure, inadequate mesh placement, erosion/migration, pain and mesh infection

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Balloon may not inflate as intended. If the balloon does not inflate during laparoscopic ventral hernia repair this may result in a delay or prolongation of the procedure, inadequate mesh placement, erosion/migration, pain and mesh infection

Code information

UDI-DI: (01)00801741201967(17)240428(10) Lot Number HUGS2043 HUGT1988

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Belgium, Czech Republic, France, Germany, Ireland, Italy, Netherlands, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK, India, Hong Kong, Greece, Chile, and Colombia.