Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 92229

3 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
November 23, 2022
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Siemens Medical Solutions USA, Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

3 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 3

Luminos Agile Max (VE10, VF10, VF11)

Z-1479-2023
Recall number
Z-1479-2023
Initiated
November 23, 2022
Classification
Class II
Status
Ongoing
Quantity
704 units in the United States (1945 units worldwide)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Under certain unlikely circumstances during a fluoroscopic examination, the imaging system Fluorospot Compact might sporadically display an incorrect air kerma/air kerma rate related to the patient reference point, according to IEC 60601-2-43. There is no impact on workflow or diagnosis. The error can occur only in fluoroscopy systems with a second (overhead) X-ray tube. Siemens Healthineers became aware of one customer site where this issue was identified during clinical operation.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Radiation Control for Health and Safety Act

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Under certain unlikely circumstances during a fluoroscopic examination, the imaging system Fluorospot Compact might sporadically display an incorrect air kerma/air kerma rate related to the patient reference point, according to IEC 60601-2-43. There is no impact on workflow or diagnosis. The error can occur only in fluoroscopy systems with a second (overhead) X-ray tube. Siemens Healthineers became aware of one customer site where this issue was identified during clinical operation.

Code information

Model: 10762472

Distribution pattern

US Nationwide - Worldwide Distribution

device · product 2 of 3

Luminos dRF Max (VE10, VF10, VF11)

Z-1480-2023
Recall number
Z-1480-2023
Initiated
November 23, 2022
Classification
Class II
Status
Ongoing
Quantity
704 units in the United States (1945 units worldwide)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Under certain unlikely circumstances during a fluoroscopic examination, the imaging system Fluorospot Compact might sporadically display an incorrect air kerma/air kerma rate related to the patient reference point, according to IEC 60601-2-43. There is no impact on workflow or diagnosis. The error can occur only in fluoroscopy systems with a second (overhead) X-ray tube.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Radiation Control for Health and Safety Act

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Under certain unlikely circumstances during a fluoroscopic examination, the imaging system Fluorospot Compact might sporadically display an incorrect air kerma/air kerma rate related to the patient reference point, according to IEC 60601-2-43. There is no impact on workflow or diagnosis. The error can occur only in fluoroscopy systems with a second (overhead) X-ray tube.

Code information

Model: 10762471

Distribution pattern

US Nationwide - Worldwide Distribution

device · product 3 of 3

LUMINOS Lotus Max (VF11)

Z-1481-2023
Recall number
Z-1481-2023
Initiated
November 23, 2022
Classification
Class II
Status
Ongoing
Quantity
704 units in the United States (1945 units worldwide)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Under certain unlikely circumstances during a fluoroscopic examination, the imaging system Fluorospot Compact might sporadically display an incorrect air kerma/air kerma rate related to the patient reference point, according to IEC 60601-2-43. There is no impact on workflow or diagnosis. The error can occur only in fluoroscopy systems with a second (overhead) X-ray tube.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Radiation Control for Health and Safety Act

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Under certain unlikely circumstances during a fluoroscopic examination, the imaging system Fluorospot Compact might sporadically display an incorrect air kerma/air kerma rate related to the patient reference point, according to IEC 60601-2-43. There is no impact on workflow or diagnosis. The error can occur only in fluoroscopy systems with a second (overhead) X-ray tube.

Code information

Model: 11574100

Distribution pattern

US Nationwide - Worldwide Distribution