device · product 1 of 1
Access Hybritech p2PSA reagent, REF B03704 and REF A49752.
- Recall number
- Z-2043-2023
- Initiated
- May 17, 2023
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Beckman Coulter, Inc.
- Quantity
- 987 kits
App-derived interpretation
Reagent lot numbers include an insufficient concentration of blocking reagent.
Official device-enrichment evidence · Sourced
Nonconforming Material/Component
Inspect official wording and provenance
Reason for recall
Reagent lot numbers include an insufficient concentration of blocking reagent.
Code information
Lot #234320, exp. 7/31/2023, UDI (01)15099590211325(17)230731(11)220731(10)234320; Lot #234134, exp. 6/30/2023, UDI (01)15099590211325(17)230630(11)220630(10)234134; and Lot #234133, exp. 6/30/2023, UDI (01)15099590204112(17)230630(11)220630(10)234133.
Distribution pattern
Distribution was made to CA, CO, CT, FL, GA, IL, IN, MA, MI, NC, NE, NJ, NY, OH, SC, TN, TX, UT, and WA. There was government distribution but no military distribution. Foreign distribution was made to Australia, Austria, Belgium, Brazil, China, Czech Republic, France, Germany, Hong Kong, India, Italy, Japan, Korea, Lithuania, Macao, Mexico, Portugal, Russia, Serbia, Singapore, Slovakia, Spain, Switzerland, Taiwan, and Vietnam.