Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 92443

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 12, 2023
Product types
Device
Classifications
Class I
Statuses
Ongoing
Recalling firm wording
Draeger Medical, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Oxylog 3000 Plus emergency and transport ventilator

Z-2018-2023
Recall number
Z-2018-2023
Initiated
June 12, 2023
Classification
Class I
Status
Ongoing
Recalling firm
Draeger Medical, Inc.
Quantity
300 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Draeger has become aware of instances where the device stopped working when in use when transitioning from battery to AC operation.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Draeger has become aware of instances where the device stopped working when in use when transitioning from battery to AC operation.

Code information

Catalog No. 5704811 & 5704813; UDI-DI: 04048675398295; All Serial No.

Distribution pattern

Distributed Nationwide. OUS Distribution: Afghanistan, Algeria, Andorra, Angola, Antigua/Barbados, Argentina, Australia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bolivia, Brunei Dar-es-S, Bulgaria, Burkina-Faso, Cambodia, Canada, Cayman Islands, Chile, China, Christmas Island, Columbia, Congo, Costa Rica, Croatia, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Equatorial Guinea, Estonia, Ethiopia, Faeroe, Finland, France, French Polynesia, French Guiana, Gabon, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guinea, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Laos, Latvia, Lebanon, Libya, Liechtenstein, Lithuania, Macau, Malawi, Malaysia, Mali, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldavia, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, North Korea, Norway, Oman, Pakistan, Panama, Pap. New Guinea, Paraguay, Peru, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., Rwanda, San Marino, Saudi Arabia, Senegal, Serbia, Singapore, Sint Maarten, Slovenia, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Suriname, Sweden, Switzerland, Taiwan, Thailand, Trinidad/Tobago, Tunisia, Turkey, Turkmenistan, Uganda, Ukraine, United Kingdom, Uruguay, Utd. Arab. Emirates, Uzbekistan, Vietnam, Yemen, Zambia, and Zimbabwe.