Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 92480

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
July 31, 2023
Product types
Device
Classifications
Class I
Statuses
Ongoing
Recalling firm wording
Datascope Corp.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-UC-0800-52, 0998-00-0800-53, 0998-UC-0800-53, 0998-00-0800-55, 0998-UC-0800-55, 0998-00-0800-65

Z-2437-2023
Recall number
Z-2437-2023
Initiated
July 31, 2023
Classification
Class I
Status
Ongoing
Recalling firm
Datascope Corp.
Quantity
9175 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Users were identifying autofill failure conditions on the devices causing pump stops.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Users were identifying autofill failure conditions on the devices causing pump stops.

Code information

All Lot Numbers. Model: 0998-00-0800-31, UDI: 10607567109053; Model: 0998-UC-0800-31, UDI: N/A; Model: 0998-00-0800-32, UDI: 10607567111117; Model: 0998-00-0800-33, UDI: 10607567109008; Model: 0998-UC-0800-33, UDI: N/A; Model: 0998-00-0800-34, UDI: 10607567111940; Model: 0998-00-0800-35, UDI: 10607567109107; Model: 0998-00-0800-45, UDI: 10607567108421; Model: 0998-00-0800-52, UDI: 10607567108438; Model: 0998-UC-0800-52, UDI: N/A; Model: 0998-00-0800-53, UDI: 10607567108391; Model: 0998-UC-0800-53, UDI: N/A; Model: 0998-00-0800-55, UDI: 10607567108414; Model: 0998-UC-0800-55, UDI: N/A; Model: 0998-00-0800-65, UDI: 10607567113432; Model: 0998-00-0800-75, UDI: 10607567112312; Model: 0998-00-0800-83, UDI: 10607567108407; Model: 0998-00-0800-85, UDI: 10607567113449

Distribution pattern

Worldwide - US Nationwide distribution.

device · product 2 of 2

Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85

Z-2438-2023
Recall number
Z-2438-2023
Initiated
July 31, 2023
Classification
Class I
Status
Ongoing
Recalling firm
Datascope Corp.
Quantity
9175 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Users were identifying autofill failure conditions on the devices causing pump stops.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Users were identifying autofill failure conditions on the devices causing pump stops.

Code information

All Lot Numbers. Model: 0998-00-0800-75, UDI: 10607567112312; Model: 0998-00-0800-83, UDI: 10607567108407; Model: 0998-00-0800-85, UDI: 10607567113449

Distribution pattern

Worldwide - US Nationwide distribution.