Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 92484

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 02, 2023
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Beckman Coulter Biomedical GmbH

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Beckman Coulter DxA Automation System, DxA 5000, REF: B87352, IVD, 200240V, 16A, 50/60HZ, RxOnly

Z-2144-2023
Recall number
Z-2144-2023
Initiated
June 02, 2023
Classification
Class II
Status
Ongoing
Quantity
146 systems

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software defect

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

A software defect in the DxA Automation System may cause tests to be run on patient samples in tube types not compatible with the requested test, leading to erroneous patient results

Code information

UDI-DI: (01)15099590707118, All DxA serial numbers

Distribution pattern

US: AL CA FL IL IN KY MI MN MO NE NJ NV NY OH PA SC TX WI WV OUS: Australia, Belgium, Canada, China, Croatia, Czechia, Egypt, France, Germany, Greece, Hong Kong, Ireland, Israel, Italy, Kazakhstan, Korea, Kuwait, Malaysia, Netherlands, New Zealand, Poland, Saudi Arabia, Spain, Switzerland, United Kingdom,

device · product 2 of 2

Beckman Coulter DxA Automation System, DxA 5000 Fit, REF: B87352, IVD, 200240V, 16A, 50/60HZ, RxOnly

Z-2145-2023
Recall number
Z-2145-2023
Initiated
June 02, 2023
Classification
Class II
Status
Ongoing
Quantity
15

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software defect

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

A software defect in the DxA Automation System may cause tests to be run on patient samples in tube types not compatible with the requested test, leading to erroneous patient results

Code information

UDI-DI: (01)15099590707118, All DxA serial numbers

Distribution pattern

US: AL CA FL IL IN KY MI MN MO NE NJ NV NY OH PA SC TX WI WV OUS: Australia, Belgium, Canada, China, Croatia, Czechia, Egypt, France, Germany, Greece, Hong Kong, Ireland, Israel, Italy, Kazakhstan, Korea, Kuwait, Malaysia, Netherlands, New Zealand, Poland, Saudi Arabia, Spain, Switzerland, United Kingdom,