Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 92615

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 24, 2023
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
NeuMoDx Molecular Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

NeuMoDx Cartridge-used for extraction, purification, amplification and detection of nucleic acids on the NeuMoDx 288 and NeuMoDx 96 Molecular Systems (NeuMoDx System(s) Ref: 100100

Z-2255-2023
Recall number
Z-2255-2023
Initiated
May 24, 2023
Classification
Class II
Status
Ongoing
Recalling firm
NeuMoDx Molecular Inc
Quantity
20,016 cartridges

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Top label on some of the NeuMoDx cartridges within these lots may have been incorrectly applied during the manufacturing process, resulting in encroachment into the PCR region. The misplaced label may interfere with the module s optics performance, thereby increasing the risk for false positive results

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Employee error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Top label on some of the NeuMoDx cartridges within these lots may have been incorrectly applied during the manufacturing process, resulting in encroachment into the PCR region. The misplaced label may interfere with the module s optics performance, thereby increasing the risk for false positive results

Code information

GTIN: 10814278020274 Lot Numbers: 117514, 117524, and 117528

Distribution pattern

US Nationwide Distribution to states of: Kansas, Florida, Texas, Ohio, New York, Michigan, Pennsylvania; and OUS (Foreign) to countries of: AT, BE, CH, DE, ES, FI, FR, GB, IT, LV, PL, RO, SE