Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 92633

5 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 26, 2023
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Olympus Corporation of the Americas

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

5 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 5

VISERA Tracheal Intubation Videoscope OLYMPUS LF Type V- For airway management which includes endoscopic observation to access airway anatomy, endotracheal/endobronchial intubation and management.

Z-2340-2023
Recall number
Z-2340-2023
Initiated
May 26, 2023
Classification
Class II
Status
Ongoing
Quantity
1433 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V and LF-P as compatible with the OER 2. Change conditions for ETO gas sterilization from 12% EtO to 100% EtO 3. Removing the reusable cleaning brush for the manual cleaning. Single Use brushes remain compatible

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling Change Control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V and LF-P as compatible with the OER 2. Change conditions for ETO gas sterilization from 12% EtO to 100% EtO 3. Removing the reusable cleaning brush for the manual cleaning. Single Use brushes remain compatible

Code information

UDI-DI: 04953170340246 All serial numbers

Distribution pattern

Nationwide

device · product 2 of 5

Tracheal Intubation Fiberscope OLYMPUS LF-P- To be used with an Olympus Light Source, documentation equipment, display monitor and other ancillary equipment for management of Endotracheal Tube

Z-2341-2023
Recall number
Z-2341-2023
Initiated
May 26, 2023
Classification
Class II
Status
Ongoing
Quantity
686 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V and LF-P as compatible with the OER 2. Change conditions for ETO gas sterilization from 12% EtO to 100% EtO 3. Removing the reusable cleaning brush for the manual cleaning. Single Use brushes remain compatible

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling Change Control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V and LF-P as compatible with the OER 2. Change conditions for ETO gas sterilization from 12% EtO to 100% EtO 3. Removing the reusable cleaning brush for the manual cleaning. Single Use brushes remain compatible

Code information

UDI-DI: 04953170051098 All serial numbers

Distribution pattern

Nationwide

device · product 3 of 5

Olympus Endoscope Reprocessor OER-Pro-For use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes and their accessories.

Z-2342-2023
Recall number
Z-2342-2023
Initiated
May 26, 2023
Classification
Class II
Status
Ongoing
Quantity
4244 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V and LF-P as compatible with the OER 2. Change conditions for ETO gas sterilization from 12% EtO to 100% EtO 3. Removing the reusable cleaning brush for the manual cleaning. Single Use brushes remain compatible

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling Change Control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V and LF-P as compatible with the OER 2. Change conditions for ETO gas sterilization from 12% EtO to 100% EtO 3. Removing the reusable cleaning brush for the manual cleaning. Single Use brushes remain compatible

Code information

UDI-DI: 04953170258589 All serial numbers

Distribution pattern

Nationwide

device · product 4 of 5

Olympus Endoscope Reprocessor OER-Mini-For use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes and their accessories.

Z-2343-2023
Recall number
Z-2343-2023
Initiated
May 26, 2023
Classification
Class II
Status
Ongoing
Quantity
241 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V and LF-P as compatible with the OER 2. Change conditions for ETO gas sterilization from 12% EtO to 100% EtO 3. Removing the reusable cleaning brush for the manual cleaning. Single Use brushes remain compatible

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling Change Control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V and LF-P as compatible with the OER 2. Change conditions for ETO gas sterilization from 12% EtO to 100% EtO 3. Removing the reusable cleaning brush for the manual cleaning. Single Use brushes remain compatible

Code information

UDI-DI:- 04953170331619 Serial Numbers: 2300314 and the last three digits less than 314

Distribution pattern

Nationwide

device · product 5 of 5

Olympus Endoscope Reprocessor OER-Elite-For use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes and their accessories.

Z-2344-2023
Recall number
Z-2344-2023
Initiated
May 26, 2023
Classification
Class II
Status
Ongoing
Quantity
1063 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V and LF-P as compatible with the OER 2. Change conditions for ETO gas sterilization from 12% EtO to 100% EtO 3. Removing the reusable cleaning brush for the manual cleaning. Single Use brushes remain compatible

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling Change Control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V and LF-P as compatible with the OER 2. Change conditions for ETO gas sterilization from 12% EtO to 100% EtO 3. Removing the reusable cleaning brush for the manual cleaning. Single Use brushes remain compatible

Code information

UDI-DI: 04953170404047 Serial Numbers: 2201375 and the last 4 digits less than 1375

Distribution pattern

Nationwide