Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 92710

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 21, 2023
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Randox Laboratories Ltd.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Randox Liquid Urine Control Level 2- intended for in vitro diagnostic use, in the quality control of Amylase, Calcium, Chloride, Cortisol, Creatinine, Glucose, hCG pregnancy, Magnesium, Microalbumin, Osmolality, pH, Phosphate Inorganic, Potassium, Protein Total, Sodium, Specific Gravity, Urea and Uric Acid (Urate) on clinical chemistry systems Catalog Number: UC5074

Z-2446-2023
Recall number
Z-2446-2023
Initiated
June 21, 2023
Classification
Class II
Status
Ongoing
Recalling firm
Randox Laboratories Ltd.
Quantity
39 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
(1) There is vial to vial variation resulting in some vials recovering positive for hCG, which should be negative and (2) high and outside range for cortisol. (3) There has been a transcription error for Creatinine in the Instructions For Use (IFU). The target and ranges for Creatinine for the Roche Creatinine Plus method have been listed incorrectly. These failures cause a delay in patient results for the above mentioned analytes.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

(1) There is vial to vial variation resulting in some vials recovering positive for hCG, which should be negative and (2) high and outside range for cortisol. (3) There has been a transcription error for Creatinine in the Instructions For Use (IFU). The target and ranges for Creatinine for the Roche Creatinine Plus method have been listed incorrectly. These failures cause a delay in patient results for the above mentioned analytes.

Code information

GTIN: 05055273207569 Batch/Lot Number: 1209UC Expiry Date: 28 Mar 24

Distribution pattern

Worldwide distribution - US Nationwide distribution in the states of AR, FL, GA, OH, VA including PR and WW distribution.