Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 92784

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
July 27, 2023
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Aomori Olympus Co., Ltd.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Olympus Distal Attachment (in Olympus Disposable EMR Kit)- Designed to be attached to the distal end of the endoscope. Disposable EMR Kit: K-001, K-002, K-003, K-004, K-011, K-012

Z-2522-2023
Recall number
Z-2522-2023
Initiated
July 27, 2023
Classification
Class II
Status
Ongoing
Recalling firm
Aomori Olympus Co., Ltd.
Quantity
1,048 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The Distal Attachment contain di-2-ethylhexyl phthalate (DEHP), which exceeds the permissible exposure level calculated according to ISO 10993-17: 2002

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The Distal Attachment contain di-2-ethylhexyl phthalate (DEHP), which exceeds the permissible exposure level calculated according to ISO 10993-17: 2002

Code information

(1)K-001: Lot Numbers: 5835330 0XI, 0YI, 12I, 14I, 15I, 16I, 17I, 26K, 27K, 28K, 29K, 2XK, 2YK, 2ZK, 31K, 32K. UDI: 4953170048562 (2) K-002 Lot Numbers: 5835430 0XI, 0YI, 12I, 13I, 14I, 16I, 17I, 27K, 2XK, 2YK, 2ZK 4953170048579 K-003 5835530 0XI, 0YI, 12I, 13I, 14I, 15I, 16I, 17I, 25K, 26K, 27K, 28K, 29K, 2XK, 2YK, 2ZK UDI:4953170048586 (3)K-004 Lot Numbers: 5835630 16I, 17I, 26K UDI: 4953170048593 (4)K-011 Lot Numbers: N2485230 0XI, 15I, 17I UDI: 4953170214035 (5) K-012 Lot Numbers: N2485330 16I, 17I UDI: 4953170214042

Distribution pattern

US Nationwide distribution.

device · product 2 of 2

Reusable distal attachments Models: MH-462, MH-463, MH-464, MH-465, MH-466, MH-483, MH-587, MH-588(US), MH-589, MH-590, MH-591, MH-592, MH-593, MH-594, MH-595, MH-596, MH-597, MH-598, MAJ-289, MAJ-290, MAJ-291, MAJ-292, MAJ-293, MAJ-294, MAJ-295, MAJ-296, MAJ-297, MAJ-663, MAJ-664, MAJ-665, MAJ-666

Z-2523-2023
Recall number
Z-2523-2023
Initiated
July 27, 2023
Classification
Class II
Status
Ongoing
Recalling firm
Aomori Olympus Co., Ltd.
Quantity
1,252 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The Distal Attachment contain di-2-ethylhexyl phthalate (DEHP), which exceeds the permissible exposure level calculated according to ISO 10993-17: 2002

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The Distal Attachment contain di-2-ethylhexyl phthalate (DEHP), which exceeds the permissible exposure level calculated according to ISO 10993-17: 2002

Code information

Model: UDI: MH-462 4953170047732 MH-463 4953170047756 MH-464 4953170047770 MH-465 4953170047794 MH-466 4953170047817 MH-483 N/A MH-587 4953170047855 MH-588 4953170364518 (US) MH-589 4953170047893 MH-590 N/A MH-591 4953170047930 MH-592 N/A MH-593 4953170047978 MH-594 4953170047992 MH-595 4953170048012 MH-596 N/A MH-597 4953170048050 MH-598 N/A MAJ-289 N/A MAJ-290 N/A MAJ-291 N/A MAJ-292 N/A MAJ-293 N/A MAJ-294 N/A MAJ-295 N/A MAJ-296 N/A MAJ-297 N/A MAJ-663 N/A MAJ-664 N/A MAJ-665 N/A MAJ-666 N/A DIstal Attachment- All lots

Distribution pattern

US Nationwide distribution.