openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
device · product 1 of 1
ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP), Material Number 11319151, in vitro diagnostic use in the quantitative determination of total protein in human urine and cerebrospinal fluid
There is a potential for ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Creatinine_2 (ECRE_2). Falsely depressed ECRE_2 results may be observed when the assay is processed after the UCFP test on ADVIA Chemistry systems.
These labels are deterministic app interpretations, not FDA categories.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
There is a potential for ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Creatinine_2 (ECRE_2). Falsely depressed ECRE_2 results may be observed when the assay is processed after the UCFP test on ADVIA Chemistry systems.
Code information
UDI-DI: 00630414279176; Lot Numbers: All lots
Distribution pattern
Worldwide - US Nationwide distribution in the states of CA, CO, CT, FL, GA, IA, MD, MI, NJ, NM, NY, OH, PA, SD, TX, UT, and VA. Global distribution.