Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 92926

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 22, 2023
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Apellis Pharmaceuticals, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Apellis Injection Kit 29g Injection Needle -single use intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin and either an 18-gauge or a 19-gauge 5 micron filter needle. Ref: IVT-KIT-29G

Z-2653-2023
Recall number
Z-2653-2023
Initiated
August 22, 2023
Classification
Class II
Status
Ongoing
Quantity
61,943 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
19-gauge x 1" inch filter needles showed appearances of a small fraction of non-preferred characteristics. A causal relationship has not been established between the structural variations in this 19-gauge filter needle and the rare events of retinal vasculitis

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

19-gauge x 1" inch filter needles showed appearances of a small fraction of non-preferred characteristics. A causal relationship has not been established between the structural variations in this 19-gauge filter needle and the rare events of retinal vasculitis

Code information

UDI-DI: 00860008043672 Kit Lots: 223186 230036 230056 230236 230316 230326 230336 230406 231926 231946 231956 231986

Distribution pattern

US Nationwide distribution.