device · product 1 of 1
therascreen ¿KRAS RGQ PCR Kit (24)-IVD qualitative PCR assay used on the Rotor-Gene Q MDx instrument for the detection of seven somatic mutations in the human KRAS oncogene REF 870021 (US IVD)
Handbook Revision (Rev13) a dilution step is indicated as an Action when having the Flag SAMPLE_INT_CTRL_FAIL. A dilution of CRC tissue could lead to a false negative result. To remove ambiguity, an alteration must be made to the S the sample result with this flag should be interpreted as invalid
These labels are deterministic app interpretations, not FDA categories.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Handbook Revision (Rev13) a dilution step is indicated as an Action when having the Flag SAMPLE_INT_CTRL_FAIL. A dilution of CRC tissue could lead to a false negative result. To remove ambiguity, an alteration must be made to the S the sample result with this flag should be interpreted as invalid
Code information
GTIN number: 04053228002048 All lots that have an expiry date of March 2023 or later
Distribution pattern
Worldwide - US Nationwide distribution in the states of AZ, CA, FL, IL, IN, KS, MA, NC,NY,OR, PA, TN, TX and the countries of KR, CN