Cardiohelp-i, blood oxygenation and carbon dioxide removal system, Part Numbers 701072780 (Cardiohelp-i), 701068550 (Slide Rail Holder)
It was found that the fixation (weld) employed by the supplier of the nut bolt to the slide rail plate was insufficient. The deficient weld allows the nut bolt to break off from the slide rail plate with slight manual force.
These labels are deterministic app interpretations, not FDA categories.
It was found that the fixation (weld) employed by the supplier of the nut bolt to the slide rail plate was insufficient. The deficient weld allows the nut bolt to break off from the slide rail plate with slight manual force.
Code information
UDI-DI: 04058863074863; Serial Numbers: 90414355 through 90414868