openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
device · product 1 of 1
Venue Go Standard Carts Ref: (H45181VC and H45103VCW), used with Venue Go R2, R3, R4 systems R2, R3, and R4 systems; pulsed doppler imaging system
Some Venue Go Standard Carts can develop an internal failure of the vertical/horizontal tilt adjustment mechanism which can result in the Venue Go system detaching from the cart and falling.
These labels are deterministic app interpretations, not FDA categories.
Some Venue Go Standard Carts can develop an internal failure of the vertical/horizontal tilt adjustment mechanism which can result in the Venue Go system detaching from the cart and falling.
Code information
Venue Go Standard Carts, Ref: (H45181VC and H45103VCW) used with Venue Go R2, R3, R4 systems R2, R3, and R4 systems (GTIN 0084068213884, 00840682190503, and 00195278516510) that were manufactured 2022-08 and before
Distribution pattern
Worldwide distribution - US Nationwide distribution including government and military distribution as well as PR.