Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 93206

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
September 06, 2023
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Beckman Coulter Biomedical Division

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

AU/DxC AU Chemistry Transferrin REF: OSR6152 Immune complexes formed in solution scatter light in proportion to their size, shape and concentration. Turbidimeters measure the reduction of incident light due to reflection, absorption, or scatter. In the procedure, the measurement of the decrease in light transmitted (increase in absorbance) through particles suspended in solution as a result of complexes formed during the antigen-antibody reaction, is the basis of this assay.

Z-0309-2024
Recall number
Z-0309-2024
Initiated
September 06, 2023
Classification
Class II
Status
Ongoing
Quantity
2,754

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The firm has identified that the Transferrin reagent does not meet the Instructions for Use (IFU) reagent on-board stability claims, which may cause control failures and/or erroneous patient samples results. The Complaint handling unit has received a total of 2 safety cases (CASE-2023-02027212 and CASE-2023-02045278) and 99 quality cases with the failure mode attributed to Transferrin reagent (OSR6152 / lot 2573) associated with reagent blank shifts for the Reagent 2 (R2), imprecision, calibration failures, erratic Quality Control (QC) recoveries, and/or erroneous high/low (serum, plasma, urine) patient results. The issue was confirmed through internal testing by Beckman Coulter on June 30, 2023. Reagent Blank shifts for R2, imprecision, calibration failures, erratic Quality Control (QC) recoveries, and/or erroneous high/low (serum, plasma, urine) patient results. The issue was confirmed through internal testing by Beckman Coulter on June 30, 2023.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The firm has identified that the Transferrin reagent does not meet the Instructions for Use (IFU) reagent on-board stability claims, which may cause control failures and/or erroneous patient samples results. The Complaint handling unit has received a total of 2 safety cases (CASE-2023-02027212 and CASE-2023-02045278) and 99 quality cases with the failure mode attributed to Transferrin reagent (OSR6152 / lot 2573) associated with reagent blank shifts for the Reagent 2 (R2), imprecision, calibration failures, erratic Quality Control (QC) recoveries, and/or erroneous high/low (serum, plasma, urine) patient results. The issue was confirmed through internal testing by Beckman Coulter on June 30, 2023. Reagent Blank shifts for R2, imprecision, calibration failures, erratic Quality Control (QC) recoveries, and/or erroneous high/low (serum, plasma, urine) patient results. The issue was confirmed through internal testing by Beckman Coulter on June 30, 2023.

Code information

UDI-DI: 15099590010492, LOT: 2573, Expiration: 01 August 2024

Distribution pattern

Worldwide - US Nationwide distribution including in the states of NY, CA, MS, GA, PA, SC, WA, TN, AL, AK, OR, MO, KS, FL, LA, OH, TX, NJ, MI, WV, MD, IL, RI, VA, NC, IN, OK, NV, NE, WY, KY, WI, ND, MA, NH, UT, CO, GU, PR, AR, AZ, ME and the countries of Canada, Netherlands, United Kingdom, Sweden, Germany, France, Switzerland, Tunisia, Australia, Belgium, Costa Rica, Uruguay, Turkey, South Africa, New Zealand, Italy, Thailand, Ghana, Russia, Hungary, Czech Republic, Slovakia, India, Austria, Poland, Romania, Ireland, Spain, Portugal, Mexico, Vietnam, Korea, Mozambique, Kazakhstan, China, Malaysia, Singapore, Kenya, Brazil, Namibia, Egypt, Panama.