Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 93378

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 25, 2023
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Ortho-Clinical Diagnostics, INc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

VITROS Chemistry Products ECO2 Slides-For in vitro diagnostic use only. VITROS Chemistry Products ECO2 Slides quantitatively measure total carbon dioxide (CO2) concentration in serum and plasma using VITROS 250/350/5,1 FS/4600 XT 3400 Chemistry Systems and the VITROS 5600/ XT 7600 Integrated System. Bicarbonate/carbon dioxide measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance. Product Code: 8262396

Z-0510-2024
Recall number
Z-0510-2024
Initiated
October 25, 2023
Classification
Class II
Status
Ongoing
Quantity
622 units 300 Slides (5 cartridge containing 60 slides)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Ortho Clinical Diagnostics (QuidelOrtho) confirmed during routine testing that some lots from Coating 3598 of VITROS ECO2 slides included carts which may contain consecutive slides that may not function as intended. These affected slides (potentially intermittent, consecutive slides within a cart or across consecutive carts) have an abnormality which may cause positively or negatively biased total carbon dioxide (CO2) results leading to either overdiagnosis of acidosis or missing alkalosis, both which may prompt unnecessary therapies such as IV fluids and/or bicarbonate or overdiagnosis of alkalosis which may also result in unnecessary use of IV fluids and electrolyte replacement or failure to detect acidosis and necessary treatment.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Ortho Clinical Diagnostics (QuidelOrtho) confirmed during routine testing that some lots from Coating 3598 of VITROS ECO2 slides included carts which may contain consecutive slides that may not function as intended. These affected slides (potentially intermittent, consecutive slides within a cart or across consecutive carts) have an abnormality which may cause positively or negatively biased total carbon dioxide (CO2) results leading to either overdiagnosis of acidosis or missing alkalosis, both which may prompt unnecessary therapies such as IV fluids and/or bicarbonate or overdiagnosis of alkalosis which may also result in unnecessary use of IV fluids and electrolyte replacement or failure to detect acidosis and necessary treatment.

Code information

UDI-DI: 10758750004645 Lot Numbers/Exp. Date: 5732-3598-4649 01-Feb-2025; 5732-3598-4650 01-Feb-2025; 5732-3598-4651 01-Feb-2025; 5732-3598-5102 01-Mar-2025; 5732-3598-5105 01-Mar-2025; 5732-3598-5189 01-Mar-2025

Distribution pattern

Nationwide Foreign: Bermuda Canada