Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 93599

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 05, 2023
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Ambu Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Ambu aView 2 A Monitor, Catalog numbers 405011000 and 405011000US2

Z-0714-2024
Recall number
Z-0714-2024
Initiated
December 05, 2023
Classification
Class II
Status
Ongoing
Recalling firm
Ambu Inc.
Quantity
11845 distributed US

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Instructions for Use have been updated to add the following warning: Use only M4 screws with the length of 14 16 mm when mounting Ambu aView 2 Advance on a VESA interface. Using longer screw lengths will penetrate the lithium-ion battery and result in a fire hazard and battery leakage which can cause severe burns, smoke inhalation and skin irritation. Using shorter screw lengths could result in unsecure device fastening.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Instructions for Use have been updated to add the following warning: Use only M4 screws with the length of 14 16 mm when mounting Ambu aView 2 Advance on a VESA interface. Using longer screw lengths will penetrate the lithium-ion battery and result in a fire hazard and battery leakage which can cause severe burns, smoke inhalation and skin irritation. Using shorter screw lengths could result in unsecure device fastening.

Code information

Catalog No. 405011000 - UDI-DI 5707480145089 Catalog No. 405011000US2 - UDI-DI 5707480156757 All products distributed since launch (May 2020)

Distribution pattern

Domestic distribution nationwide.