Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 93636

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 26, 2024
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
BioFire Diagnostics, LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

BIOFIRE Blood Culture Identification 2 (BCID2) Panel when used with: BD BACTEC Lytic Anaerobic medium BD BACTEC Peds Plus medium BD BACTEC Plus Aerobic medium BD BACTEC Plus Anaerobic medium BD BACTEC Standard Aerobic medium BD BACTEC Standard Anaerobic medium The BIOFIRE Blood Culture Identification 2 (BCID2) Panel is a multiplex nucleic acid assay for identification of microorganisms and resistance markers from positive blood cultures.

Z-1139-2024
Recall number
Z-1139-2024
Initiated
January 26, 2024
Classification
Class II
Status
Ongoing
Recalling firm
BioFire Diagnostics, LLC
Quantity
32,910 panels/pouches

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Due to a trend in false positive Candida tropicalis results on blood culture panel when using certain BD BACTEC vial mediums

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Due to a trend in false positive Candida tropicalis results on blood culture panel when using certain BD BACTEC vial mediums

Code information

For BioFire Blood Culture Identification 2 (BDID2) Panel Part Number: RFIT-ASY-0147 UDI-DI code: 00815381020338 Lot Numbers: All non-expired BCID2 Panel lots, if used with the BD BACTEC vials including but not limited to, the bottle types listed in Table 1. NOTE: The affected blood culture vial types most likely contain non-viable Candida tropicalis nucleic acid/DNA fragments that can be detected by the BCID2 Panel. Therefore, all lots of the BioFire BCID2 Panel would be affected if they are used with the affected blood culture vials. Neither the BioFire BCID2 test kits nor the BD BACTEC vials have a defect - both products are performing as intended.

Distribution pattern

Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Andorra, United Arab Emirates, Armenia, Argentina, Austria, Australia, Bosnia, Belgium, Bulgaria, Bahrain, Brunei, Bolivia, Brazil, Canada, Switzerland, Cote D'Ivoire, Chile, China, Colombia, Costa Rica, Cyprus, Cech Republic, Germany, Denmark, Dominican Republic, Algeria, Ecuador, Estonia, Egypt, Spain, Finland, France, Gabon, United Kingdom, Georgia, Guadeloupe, Greece, Guatemala, Hong Kong, Honduras, Croatia, Hungary, Iceland, Indonesia, Ireland, Israel, India, Italy, Jordan, Japan, Korea, Kuwait, Lebanon, Lithuania, Latvia, Morocco, Macedonia, Mali, Mongolia, Macau, Mexico, Malaysia, New Caledonia, Nigeria, Nicaragua, Netherlands, Norway, New Zealand, Oman, Panama, Peru, French Polynesia, Philippines, Pakistan, Poland, St. Pierre/Miquelon, Portugal, Paraguay, Qatar, Reunion, Romania, Serbia, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, El Salvador, Thailand, Tunisia, Turkey, Taiwan, Uganda, Uruguay, Vatican City, Venezuela, Vietnam, South Africa, and Zimbabwe.