device · product 1 of 1
Cobalt XT DR MRI SureScan Implantable Cardioverter Defibrillator, product number DDPA2D1
During internal review of Polaris battery assembly data from the HiPot operation, a test systems engineer noticed that some batteries had gone through the High Potential test twice. This was able to occur because the test screen did not prevent retesting after a failure. Some batteries were therefore retested and passed.
These labels are deterministic app interpretations, not FDA categories.
During internal review of Polaris battery assembly data from the HiPot operation, a test systems engineer noticed that some batteries had gone through the High Potential test twice. This was able to occur because the test screen did not prevent retesting after a failure. Some batteries were therefore retested and passed.
Code information
UDI/DI 00763000178338, Serial Number RSL604096S
Distribution pattern
US Nationwide distribution in the state of Florida.