device · product 1 of 1
The Sofia 2 Flu + SARS Antigen FIA employs immunofluorescence technology in a sandwich design that is used with Sofia 2 to detect nucleocapsid protein from influenza A, influenza B, and SARS-CoV-2. This test allows for the detection of SARS-CoV and SARS-CoV-2, but does not differentiate between the two viruses.
- Recall number
- Z-1086-2024
- Initiated
- January 03, 2024
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Quidel Corporation
- Quantity
- 312 Kits
App-derived interpretation
One lot of the Sofia 2 Flu + SARS Antigen FIA (Lot 709083) was labelled with incorrect expiration dates. The kits contain pouched cassettes labeled with an expiration date of May 31, 2025, however the correct expiration date for the pouches is September 12, 2024. The kits are labeled with an expiration date of April 30, 2025, however, the correct expiration date for the kits is September 12, 2024. The issue was discovered on November 28, 2023 by a supply chain employee who determined that the kits had expiration dates beyond the FDA authorized shelf life.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Inspect official wording and provenance
Reason for recall
One lot of the Sofia 2 Flu + SARS Antigen FIA (Lot 709083) was labelled with incorrect expiration dates. The kits contain pouched cassettes labeled with an expiration date of May 31, 2025, however the correct expiration date for the pouches is September 12, 2024. The kits are labeled with an expiration date of April 30, 2025, however, the correct expiration date for the kits is September 12, 2024. The issue was discovered on November 28, 2023 by a supply chain employee who determined that the kits had expiration dates beyond the FDA authorized shelf life.
Code information
UDI-DI: 30014613339229. Lot: 709083
Distribution pattern
US Nationwide distribution in the states of FL, TN, IL, LA, NY, IN.