device · product 1 of 1
Access Total T4, REF 33800, IVD, thyroxine reagent assay.
- Recall number
- Z-1303-2024
- Initiated
- February 08, 2024
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Beckman Coulter, Inc.
- Quantity
- 64,009 kits
App-derived interpretation
Four lots of Access Total T4 reagent lots demonstrated imprecision outside of the 10% specification which may result in erroneous or delayed results.
Official device-enrichment evidence · Sourced
Vendor change control
Inspect official wording and provenance
Reason for recall
Four lots of Access Total T4 reagent lots demonstrated imprecision outside of the 10% specification which may result in erroneous or delayed results.
Code information
Lot numbers: 338555, exp. 31 JUL 2024; 338556, exp. 31 AUG 2024; 338734, exp. 30 SEP 2024; and 338958, exp. 31 OCT 2024; UDI-DI 15099590225742.
Distribution pattern
Worldwide distribution - US Nationwide, including Puerto Rico and the countries of Bahrain, Bangladesh, Bolivia, Brazil, Canada, China, Croatia, Czech Republic, Denmark, Egypt, El Salvador, Georgia, Greece, Hong Kong, India, Kazakhstan, Mexico, Namibia, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Spain, Taiwan, Thailand, Turkey, U.A.E, Vietnam, and Zimbabwe.