Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 94067

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 16, 2024
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Olympus Corporation of the Americas

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

InstaClear Lens Cleaner-indicated for use during routine diagnostic procedures and during endoscopic sinus surgery. Model Number: LCS1500UNOL LCS1800UNST LCS4K00UNOL LCS1530BTOL LCS1830BTST LCS4K30BTOL LCS1530TPOL LCS1830TPST LCS4K30TPOL LCS1545BTOL LCS1845BTST LCS4K45BTOL LCS1545TPOL LCS1845TPST LCS4K45TPOL LCS1570BTOL LCS1870BTST LCS4K70BTOL LCS1570TPOL LCS1870TPST LCS4K70TPOL

Z-1368-2024
Recall number
Z-1368-2024
Initiated
January 16, 2024
Classification
Class II
Status
Ongoing
Quantity
102,279 units (OUS Only)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Japanese IFU version had slightly different language and illustrations from the English IFU, which could contribute to the tip and plate falling off. In order to maintain consistency globally, the Japanese IFU will be updated to reflect the instructions in the English IFU

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Japanese IFU version had slightly different language and illustrations from the English IFU, which could contribute to the tip and plate falling off. In order to maintain consistency globally, the Japanese IFU will be updated to reflect the instructions in the English IFU

Code information

UDI-DI: N/A All Lot Numbers

Distribution pattern

International distribution to the country of Japan.