openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
device · product 1 of 1
Assembly, Array Guide TMINI, REF: 109101, part of the TMINI Knee Instrument Set, used with the TMINI Miniature Robotic System
There is a galling effect (cold-weld) caused by friction between the array fixation pins and the TMINI array guide holes due to insufficient clearance between the outer diameter of the fixation pins and the inner diameter of the TMINI array guide holes. As a result, the array guide has the possibility of binding to the array fixation pins during drilling.
These labels are deterministic app interpretations, not FDA categories.
There is a galling effect (cold-weld) caused by friction between the array fixation pins and the TMINI array guide holes due to insufficient clearance between the outer diameter of the fixation pins and the inner diameter of the TMINI array guide holes. As a result, the array guide has the possibility of binding to the array fixation pins during drilling.
Code information
UDI-DI: M9461091010. Lot Numbers: 2316701, 2232601, 23030, 2228501, 2327802, 22152
Distribution pattern
US Nationwide distribution in the states of AL, OH, NV, TX, UT, CA, MO.