openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
device · product 1 of 4
Baxter Peri-Guard Repair Patch (legacy product codes): a) PC0404NBIO, b) PC0608NBIO, c) PC0814NBIO, d) PC1016NBIO
A new contraindication was added to the IFU which states that Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices and Supple Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices.
These labels are deterministic app interpretations, not FDA categories.
A new contraindication was added to the IFU which states that Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices and Supple Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices.
Code information
a) PC0404NBIO, UDI/DI 00085412530529; b) PC0608NBIO, UDI/DI 00085412530567; c) PC0814NBIO, UDI/DI 00085412530581; d) PC1016NBIO, UDI/DI 00085412530604; ALL LOTS WITHIN EXPIRY
Distribution pattern
US Distribution to states of: CT, VA, TX, PA and IL; and OUS (foreign) countries of: Austria, Finland, Germany, Greece, Ireland, Italy, Poland, Spain, Sweden, Switzerland, and United Kingdom.
device · product 2 of 4
Baxter Supple Peri-Guard Repair Patch (legacy product codes): a) PC0404SNBIO, b) PC0608NSBIO, c) PC0814SNBIO, d) PC1016SNBIO
A new contraindication was added to the IFU which states that Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices and Supple Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices.
These labels are deterministic app interpretations, not FDA categories.
A new contraindication was added to the IFU which states that Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices and Supple Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices.
Code information
a) PC0404SNBIO, UDI/DI 00085412530642; b) PC0608SNBIO, UDI/DI 00085412530673; c) PC0814SNBIO, UDI/DI 00085412530697; d) PC1016SNBIO, UDI/DI 00085412530710; ALL LOTS WITHIN EXPIRY
Distribution pattern
US Distribution to states of: CT, VA, TX, PA and IL; and OUS (foreign) countries of: Austria, Finland, Germany, Greece, Ireland, Italy, Poland, Spain, Sweden, Switzerland, and United Kingdom.
device · product 3 of 4
Baxter Peri-Guard Repair Patch (new product codes): a) PG0404, b) PG0608, c) PG0814, d) PG1016, e) PG1225
A new contraindication was added to the IFU which states that Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices and Supple Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices.
These labels are deterministic app interpretations, not FDA categories.
A new contraindication was added to the IFU which states that Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices and Supple Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices.
Code information
a) PG0404, UDI/DI 00085412818405; b) PG0608, UDI/DI 00085412818443; c) PG0814, UDI/DI 00085412818597; d) PG1016, UDI/DI 00085412818610; e) PG1225, UDI/DI 00085412818658; ALL LOTS WITHIN EXPIRY
Distribution pattern
US Distribution to states of: CT, VA, TX, PA and IL; and OUS (foreign) countries of: Austria, Finland, Germany, Greece, Ireland, Italy, Poland, Spain, Sweden, Switzerland, and United Kingdom.
device · product 4 of 4
Baxter Supple Peri-Guard Repair Patch (new product codes): a) SPG0404, b) SPG0406, c) SPG0608, d) SPG0814, e) SPG1016
A new contraindication was added to the IFU which states that Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices and Supple Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices.
These labels are deterministic app interpretations, not FDA categories.
A new contraindication was added to the IFU which states that Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices and Supple Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices.
Code information
a) SPG0404, UDI/DI 00085412818672; b) SPG0406, UDI/DI 00085412818696; c) SPG0608, UDI/DI 00085412818313; d) SPG0814, UDI/DI 00085412818375; e) SPG1016, UDI/DI 00085412818429; ALL LOTS WITHIN EXPIRY
Distribution pattern
US Distribution to states of: CT, VA, TX, PA and IL; and OUS (foreign) countries of: Austria, Finland, Germany, Greece, Ireland, Italy, Poland, Spain, Sweden, Switzerland, and United Kingdom.