Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 94257

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
March 15, 2024
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
SPINEART SA

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

PERLA TL 25D SCREW REF: TLF-DS 45 25-S TLF-DS 45 30-S TLF-DS 45 35-S TLF-DS 45 40-S TLF-DS 45 45-S TLF-DS 55 30-S TLF-DS 55 35-S TLF-DS 55 40-S TLF-DS 55 45-S TLF-DS 65 35-S TLF-DS 65 40-S TLF-DS 65 45-S TLF-DS 65 50-S The PERLA¿ TL system is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (painful degeneration of the disc), spondylolisthesis, trauma, spinal stenosis, deformities (i.e. scoliosis, kyphosis, or lordosis), tumor and failed previous fusion (pseudarthrosis).

Z-1758-2024
Recall number
Z-1758-2024
Initiated
March 15, 2024
Classification
Class II
Status
Ongoing
Recalling firm
SPINEART SA
Quantity
242 systems

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Due to receiving information regarding difficulties with the insertion of a setscrew inside a screw head. Issues involved, either: 1. The setscrew is not tightened to torque 2. The setscrew is tightened to torque but position askew

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Due to receiving information regarding difficulties with the insertion of a setscrew inside a screw head. Issues involved, either: 1. The setscrew is not tightened to torque 2. The setscrew is tightened to torque but position askew

Code information

Model Numbers/Lot-Batch Numbers: TLF-DS 45 25-S 7-4407 TLF-DS 45 25-S 7-6207 TLF-DS 45 30-S 7-4244 TLF-DS 45 30-S 7-6226 TLF-DS 45 35-S 7-4242 TLF-DS 45 35-S 7-5470 TLF-DS 45 35-S 7-6684 TLF-DS 45 40-S 7-6232 TLF-DS 45 45-S 7-5474 TLF-DS 45 45-S 7-6874 TLF-DS 55 30-S 7-4409 TLF-DS 55 30-S 7-4695 TLF-DS 55 30-S 7-4965 TLF-DS 55 35-S 7-4413 TLF-DS 55 35-S 7-6455 TLF-DS 55 40-S 7-4415 TLF-DS 55 40-S 7-4417 TLF-DS 55 45-S 7-4411 TLF-DS 55 45-S 7-5055 TLF-DS 55 45-S 7-6087 TLF-DS 55 45-S 7-6459 TLF-DS 65 35-S 7-4405 TLF-DS 65 35-S 7-6230 TLF-DS 65 35-S 7-6461 TLF-DS 65 35-S 7-6882 TLF-DS 65 40-S 7-4248 TLF-DS 65 40-S 7-4250 TLF-DS 65 40-S 7-5480 TLF-DS 65 40-S 7-6228 TLF-DS 65 40-S 7-6554 TLF-DS 65 40-S 7-6740 TLF-DS 65 45-S 7-4582 TLF-DS 65 45-S 7-5484 TLF-DS 65 45-S 7-5808 TLF-DS 65 45-S 7-6668 TLF-DS 65 45-S 7-6742 TLF-DS 65 45-S 7-7299 TLF-DS 65 50-S 7-5488 TLF-DS 65 50-S 7-9264 UDI-DI for associated Model Numbers: TLF-DS 45 25-S 07640185345799 TLF-DS 45 30-S 07640185345805 TLF-DS 45 35-S 07640185345812 TLF-DS 45 40-S 07640185345829 TLF-DS 45 45-S 07640185345836 TLF-DS 55 30-S 07640185345850 TLF-DS 55 35-S 07640185345867 TLF-DS 55 40-S 07640185345874 TLF-DS 55 45-S 07640185345881 TLF-DS 60 35-S 07640305166082

Distribution pattern

U.S. Nationwide distribution in the states of CA, and KY. Not provided due to affected devices were manufactured O.U.S.