device · product 1 of 1
Atellica CH Iron3 - IVD use in the quantitative measurement of iron in human serum and plasma (lithium heparin and sodium heparin) using the Atellica¿ CI Analyzer and Atellica¿ CH Analyzer Siemens Material Number (SMN): 11537211
- Recall number
- Z-1549-2024
- Initiated
- March 13, 2024
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 8721 units
App-derived interpretation
Potential for falsely elevated Chol_2, LDLC, and Trig_2 results on the Atellica CH and Atellica CI analyzers when the previous result in the cuvette was Iron3. Results in a positive bias ranging from 2-16% -impacts calibrator, quality control (QC), and patient results.
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
Potential for falsely elevated Chol_2, LDLC, and Trig_2 results on the Atellica CH and Atellica CI analyzers when the previous result in the cuvette was Iron3. Results in a positive bias ranging from 2-16% -impacts calibrator, quality control (QC), and patient results.
Code information
UDI-DI: 00630414610849 All lots
Distribution pattern
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Curacao, St¿Eus, Czech Republic, Denmark, Estonia, Finland, France, Germany, ¿ Greece, Hungary, India, Ireland, Israel, Italy, Latvia, Lithuania, Mexico, Netherlands, ¿¿¿¿ New Zealand, Norway, Paraguay, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, U.A.E., United¿Kingdom.¿