Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 94441

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
April 11, 2024
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
FUJIFILM Healthcare Americas Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Noblus AC Adapter used with the Noblus Ultrasound Imaging System Model : AHM250PS24

Z-1884-2024
Recall number
Z-1884-2024
Initiated
April 11, 2024
Classification
Class II
Status
Ongoing
Quantity
418 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
AC adapters of Noblus can heat up, and cause smoke in the worst case, when an excessive mechanical stress is applied to the socket for power cable (frequent insertion/removal of the power cable).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

AC adapters of Noblus can heat up, and cause smoke in the worst case, when an excessive mechanical stress is applied to the socket for power cable (frequent insertion/removal of the power cable).

Code information

UDI-DI: (01) 04580292771704 Noblus Ultrasound Imaging System AC Adapter Serial Number Before K17160001 (Excluding K17160001)

Distribution pattern

US Nationwide distribution.