Brand Name: Cerene Cryotherapy Device Product Name: DEVICE, THERMAL ABLATION, ENDOMETRIAL Model/Catalog Number: FGS-7000 Product Description: The Cerene Cryotherapy Device is an endometrial ablation device that uses nitrous oxide (N2O) to freeze and ablate the endometrium to reduce future menstrual bleeding. The device is intended for use by healthcare professionals who have received appropriate training and are familiar with the principles, clinical applications, complications, side effects, and hazards commonly associated with endometrial ablation. The Cerene Device is single use, disposable, and provided sterile. The average procedure time is 7 minutes, with 2.5 minutes of active cryoablation. Ablation is achieved throughout the uterine cavity through the use of cryothermic energy. The cryothermic energy is provided by a liquid-to-gas phase change of N2O. During the 2.5-minute treatment cycle, liquid N2O (originating from a small Cylinder located in the device handle) flows through a delivery line and into an inflow line with multiple jets. This liquid N2O is infused into an ultra-thin polyurethane Liner, where it converts into gas. The gaseous N2O is exhausted through the Exhaust Hose exiting the bottom of the handle. Component: N/A
- Recall number
- Z-1889-2024
- Initiated
- May 03, 2024
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Channel Medsystems, Inc.
- Quantity
- 1772
App-derived interpretation
Error code 003 was not listed in certain rows of the Device Instructions for Use Troubleshooting section (i.e., Table 22. Summary of LCD Messages and Error Codes).
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
Error code 003 was not listed in certain rows of the Device Instructions for Use Troubleshooting section (i.e., Table 22. Summary of LCD Messages and Error Codes).
Code information
Lot Code: Model No: FGS-7000 UDI-DI: 00850008595035 Lot Numbers: All non-expired lots manufactured through February 2024 101635693, 101635694, 101635695, 101635696, 102211255, 102211256, 102211257, 102211258, 102887363, 102887364, 103151029, 103151030, 103151031, 103151032, 103489753, 103489754, 103489755, 103489756, 103489757, 103925402, 104008714, 104008715, 104521903, 104521904, 104521905, 104521906, 104521907, 104808352, 104808353, 105418198, 105733204, 105733205
Distribution pattern
Worldwide - US Nationwide distribution including in the states of Alabama, Arizona, Arkansas, California, Colorado, Connecticut, Florida, Georgia, Hawaii, Illinois, Indiana, Kansas, Kentucky, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New York, North Carolina, Ohio, Pennsylvania, Rhode Island, South Carolina, Tennessee, Texas, Utah, Virginia, Wisconsin and the country of Netherlands.