Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 94506

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 03, 2024
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Channel Medsystems, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Brand Name: Cerene Cryotherapy Device Product Name: DEVICE, THERMAL ABLATION, ENDOMETRIAL Model/Catalog Number: FGS-7000 Product Description: The Cerene Cryotherapy Device is an endometrial ablation device that uses nitrous oxide (N2O) to freeze and ablate the endometrium to reduce future menstrual bleeding. The device is intended for use by healthcare professionals who have received appropriate training and are familiar with the principles, clinical applications, complications, side effects, and hazards commonly associated with endometrial ablation. The Cerene Device is single use, disposable, and provided sterile. The average procedure time is 7 minutes, with 2.5 minutes of active cryoablation. Ablation is achieved throughout the uterine cavity through the use of cryothermic energy. The cryothermic energy is provided by a liquid-to-gas phase change of N2O. During the 2.5-minute treatment cycle, liquid N2O (originating from a small Cylinder located in the device handle) flows through a delivery line and into an inflow line with multiple jets. This liquid N2O is infused into an ultra-thin polyurethane Liner, where it converts into gas. The gaseous N2O is exhausted through the Exhaust Hose exiting the bottom of the handle. Component: N/A

Z-1889-2024
Recall number
Z-1889-2024
Initiated
May 03, 2024
Classification
Class II
Status
Ongoing
Recalling firm
Channel Medsystems, Inc.
Quantity
1772

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Error code 003 was not listed in certain rows of the Device Instructions for Use Troubleshooting section (i.e., Table 22. Summary of LCD Messages and Error Codes).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Error code 003 was not listed in certain rows of the Device Instructions for Use Troubleshooting section (i.e., Table 22. Summary of LCD Messages and Error Codes).

Code information

Lot Code: Model No: FGS-7000 UDI-DI: 00850008595035 Lot Numbers: All non-expired lots manufactured through February 2024 101635693, 101635694, 101635695, 101635696, 102211255, 102211256, 102211257, 102211258, 102887363, 102887364, 103151029, 103151030, 103151031, 103151032, 103489753, 103489754, 103489755, 103489756, 103489757, 103925402, 104008714, 104008715, 104521903, 104521904, 104521905, 104521906, 104521907, 104808352, 104808353, 105418198, 105733204, 105733205

Distribution pattern

Worldwide - US Nationwide distribution including in the states of Alabama, Arizona, Arkansas, California, Colorado, Connecticut, Florida, Georgia, Hawaii, Illinois, Indiana, Kansas, Kentucky, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New York, North Carolina, Ohio, Pennsylvania, Rhode Island, South Carolina, Tennessee, Texas, Utah, Virginia, Wisconsin and the country of Netherlands.