Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 94557

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
April 15, 2024
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

MyoSPECT System, Model H3912AA and H3912B, and MyoSPECT ES System, Model H3912AB and H3912BD, dedicated cardiac SPECT (single photon emission computerized tomography system).

Z-2040-2024
Recall number
Z-2040-2024
Initiated
April 15, 2024
Classification
Class II
Status
Ongoing
Quantity
133 devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
There is a potential issue regarding the service handles for the internal lead (Pb) covers in that the handles could loosen or release which could cause a cover to drop potentially resulting in injury to service personnel due to the weight of the cover.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

There is a potential issue regarding the service handles for the internal lead (Pb) covers in that the handles could loosen or release which could cause a cover to drop potentially resulting in injury to service personnel due to the weight of the cover.

Code information

MyoSPECT manufactured on or before 2024-02-13: Model H3912AA, GTIN 00195278421586 - Serial numbers NGCC80094, NGCB80033, NGCB80044, NGCB80045, NGCA80017, NGCB80039, NGCA80006, NGCA80011, NGCA80003, NGCA80012, NGCB80052, NGCB80060, NGCC80074, NGCC80081, NGCB80038, NGCD80107, NGCB80070, NGCC80093, NGCD80106, NGCB80026, NGCB80050, NGCB80058, NGCC80084, NGCB80027, NGCB80072, NGCB80032, NGCB80031, NGCC80080, NGCB80067, NGCA80021, NGCA80010, NGCA80015, NGCB80043, NGCC80090, NGCC80091, NGCD80104, NGCB80028, NGCB80022, NGCB80023, NGCB80048, NGCB80035, NGCD80103, NGCC80100, NGCC80083, NGCB80042, NGCC80075, NGCB80068, NGCB80040, NGCB80051, NGCB80064, NGCC80078, NGCC80079, NGCC80087, NGCC80092, NGCD80105, NGCD80108, NGCB80059, NGCA80009, NGCC80086, NGCC80095, NGCA80020, NGCC80088, NGCB80055, NGCB80029, NGCB80041, NGCC80076, NGCA80016, NGCB80036, NGCC80098, NGCB80037, NGCB80046, NGCA80007, NGCC80082, NGCB80061, NGCA80002, NGCC80102, NGCB80057, NGCC80089, NGCB80030, NGCC80097, NGCA80005, NGCB80049, NGCC80101, NGCB80071, NGCA80001, NGCB80066, NGCB80065, NGCC80085, NGCC80077, NGCB80069, NGCA80014, MYGA10001, NGCC80096, NGCB80024, NGCB80047, NGCC80099, NGCB80062, NGCB80034, NGCA80008, NGCA80013, NGCB80025, NGCA80004, NGCA80019, NGCB80056, NGCB80063, NGCB80054, and NGCB80053. Model H3912BC, GTIN 00195278488619 - Serial number MYOC93001. MyoSPECT ES manufactured on or before 2024-02-15: Model H3912AB, GTIN 00195278421579 - Serial numbers NGEB81002, NGED81025, NGEB81007, NGEB81006, NGEC81019, NGEB81012, NGEB81008, NGEC81014, NGEB81005, NGEC81017, NGEC81022, NGEC81021, NGEC81020, NGEB81011, NGEB81010, NGEC81023, NGEC81024, NGEC81013, NGEB81009, NGEB81003, NGEB81004, NGEC81018, NGEC81015, and NGEC81016. Model H3912BD, GTIN 00195278488626 - Serial number MYEB94001.

Distribution pattern

Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, LA, MA, MD, MI, MO, NC, ND, NV, OH, OK, OR, PA, TN, TX, VA, and WA. The countries of Argentina, Belgium, Brazil, Canada, Czech Republic, Finland, France, India, Israel, Italy, Japan, Malaysia, Malta, Poland, Singapore, Spain, Taiwan, Turkey, and United Kingdom.