Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 94623

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 13, 2024
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Aesculap Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Aeos Robotic Digital Microscope, Product Code: PV010

Z-2139-2024
Recall number
Z-2139-2024
Initiated
May 13, 2024
Classification
Class II
Status
Ongoing
Recalling firm
Aesculap Inc
Quantity
19 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Possibility of the robotic arm to drop more than 10cm from its original position after pressing the emergency stop button due to failure of the integrated safety mechanism.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Possibility of the robotic arm to drop more than 10cm from its original position after pressing the emergency stop button due to failure of the integrated safety mechanism.

Code information

Product Code: PV010, UDI/DI: 04046955206742; Serial Numbers: 1011, 1005, 1057, 1032, 1038, 1054, 1036, 1060, 1056, 1089, 1093, 1027, 1061, 1004, 1026, 1035, 1097, 1084, 1063

Distribution pattern

US Nationwide distribution in the states of AR, FL, IL, IN, KY, MI, NC, NE, OK, SD, TX.