Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 94644

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 13, 2024
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Draeger Medical Systems, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Infinity Acute Care System (IACS) Monitoring System, Infinity M500 with Power, Model No. MS20407. Rev 20, 21, and 23

Z-2189-2024
Recall number
Z-2189-2024
Initiated
May 13, 2024
Classification
Class II
Status
Ongoing
Quantity
34,102 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Device is not in full compliance to Type CF requirements of IEC 60601-1 and IEC 60601-2-34 standards. Therefore, there is a potential risk of electric shock. Health consequences may include electrical shock, burns, and cardiac arrhythmia.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Device is not in full compliance to Type CF requirements of IEC 60601-1 and IEC 60601-2-34 standards. Therefore, there is a potential risk of electric shock. Health consequences may include electrical shock, burns, and cardiac arrhythmia.

Code information

Model Number: MS20407. UDI-DI: 04049098054447. Rev 20, 21, and 23

Distribution pattern

US Nationwide distribution. International distribution to Afghanistan, Argentina, Armenia, Australia, Austria, Belarus, Belgium, Bolivia, Botswana, Brazil, Brunei Dar-es-S, Bulgaria, Canada, Chile, China, Columbia, Costa Rica, Croatia, Czech Republic, Egypt, Estonia, Finland, France, French Guiana, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Kuwait, Latvia, Liechtenstein, Lithuania, Malaysia, Malta, Martinique, Mexico, Monaco, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, T¿rkiye, Turkmenistan, Ukraine, United Kingdom, Uruguay, United Arab Emirates, Vietnam.