Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 94650

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 08, 2024
Product types
Device
Classifications
Class III
Statuses
Ongoing
Recalling firm wording
Ortho-Clinical Diagnostics, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

VITROS Immunodiagnostic Products Vitamin B12 Range Verifier - In vitro diagnostic use ly. For use in verifying the calibration range of the VITROS ECi/ECiQ/3600 Immunodiagnostic Systems and the VITROS 5600/XT 7600 Integrated Systems when used for the quantitative measurement of vitamin B12 - Product Code: 6800872

Z-2019-2024
Recall number
Z-2019-2024
Initiated
May 08, 2024
Classification
Class III
Status
Ongoing
Quantity
76 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The results generated by the VITROS Immunodiagnostic Products Vitamin B12 Range Verifier Lot# 1200 RV Level 2 may cause range verification failure. The issue is detectable by the user when the product fails to meet the required range during processing of RV Level 2 of this lot. There is a potential for a delay in patient test results should the user have a need to switch to an alternate method of range verification including use of an alternate lot. As this issue affects the upper limit range verification, there is no impact to VITROS Immunodiagnostic Products Vitamin B12 assay performance. Were the assay to not perform as expected, the failure would be detectable by Quality Control. Past patient sample results are unaffected, and no review of reported results is required.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The results generated by the VITROS Immunodiagnostic Products Vitamin B12 Range Verifier Lot# 1200 RV Level 2 may cause range verification failure. The issue is detectable by the user when the product fails to meet the required range during processing of RV Level 2 of this lot. There is a potential for a delay in patient test results should the user have a need to switch to an alternate method of range verification including use of an alternate lot. As this issue affects the upper limit range verification, there is no impact to VITROS Immunodiagnostic Products Vitamin B12 assay performance. Were the assay to not perform as expected, the failure would be detectable by Quality Control. Past patient sample results are unaffected, and no review of reported results is required.

Code information

UDI-DI: 10758750000937 Lot Number: 1200 Exp. Date: 30-Sep-2024

Distribution pattern