device · product 1 of 1
US AquaA 900H-Typ USA-3 x 208v-Intended for the purification of water to be used for hemodialysis Vivonic Part Numbers: G02020110 Fresenius Part Number: 24-090A-1
Concentrate Flow was exceeding the maximum 19.9 bar set point. The standard default set point range is between 4 and 18.Error message (F020104, "concentrate pressure alarm limit exceeded") concentrate pressure is too high and stops the dialysis water supply.
These labels are deterministic app interpretations, not FDA categories.
Concentrate Flow was exceeding the maximum 19.9 bar set point. The standard default set point range is between 4 and 18.Error message (F020104, "concentrate pressure alarm limit exceeded") concentrate pressure is too high and stops the dialysis water supply.
Code information
UDI-DI: 04251285200357 Serial Numbers: 2AAV0003 2AAV0029 2AAV0037 3AAV0087
Distribution pattern
US Nationwide distribution in the states of AL, IN, AK, WI.