Event 94840

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 22, 2024
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Beckman Coulter, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.

Z-2735-2024
Recall number
Z-2735-2024
Initiated
May 22, 2024
Classification
Class II
Status
Ongoing
Recalling firm
Beckman Coulter, Inc.
Quantity
183 analyzers

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
When a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS) and is running with the system software versions 1.17.0 and below, the analyzer does not send a message to the LAS when the sample wheel is full, and the LAS continues sending samples despite the limited capacity to process new samples. Therefore, sample tests may be aliquoted but not processed when the number of tests associated with onboard patient samples exceeds the throughput limit. If this issue occurs test results are flagged with a sample wheel timeout (SWT) error flag, and there may be a delay in reporting test results which could lead to a delay in patient treatment.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

When a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS) and is running with the system software versions 1.17.0 and below, the analyzer does not send a message to the LAS when the sample wheel is full, and the LAS continues sending samples despite the limited capacity to process new samples. Therefore, sample tests may be aliquoted but not processed when the number of tests associated with onboard patient samples exceeds the throughput limit. If this issue occurs test results are flagged with a sample wheel timeout (SWT) error flag, and there may be a delay in reporting test results which could lead to a delay in patient treatment.

Code information

Software versions 1.17.0 or below, UDI-DI 15099590732103.

Distribution pattern

Worldwide - US Nationwide distribution in the states of CA, FL, GA, IN, KS, ME, NH, NJ, OK, PA, TX, and WA. The countries of Australia, Austria, Bahrain, Belgium, Brazil, Croatia, Czechia, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Switzerland, Taiwan, and United Kingdom of Great Britain and Northern Ireland.