device · product 1 of 1
Orthopantomograph OP300 dental computed tomography x-ray system. The X-ray device produces conventional 2D X-ray images and Xray projection images for the reconstruction of a 3D view. The device is operated and used by qualified healthcare professionals.
It was noticed in production on 28.10.2021, that the final testing database gives a pass result for Pan ADC repeatability test even when the test result is not within the set acceptance criteria (repeatability within 4%).
These labels are deterministic app interpretations, not FDA categories.
It was noticed in production on 28.10.2021, that the final testing database gives a pass result for Pan ADC repeatability test even when the test result is not within the set acceptance criteria (repeatability within 4%).