Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 95028

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
July 22, 2024
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
EVIDENT SCIENTIFIC INC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

SlideView VS200 slide viewer,

Z-2416-2024
Recall number
Z-2416-2024
Initiated
July 22, 2024
Classification
Class II
Status
Ongoing
Recalling firm
EVIDENT SCIENTIFIC INC
Quantity
N/A

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit and destroyed the fuse in the laser. This caused the shutter to remain in its most recent state; either in an open or closed position. Once the fuse was destroyed the shutter could not be operated.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Radiation Control for Health and Safety Act

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit and destroyed the fuse in the laser. This caused the shutter to remain in its most recent state; either in an open or closed position. Once the fuse was destroyed the shutter could not be operated.

Code information

65159 67105 67104 69153 67106 69154

Distribution pattern

US